Heathkit Model IP-28
1 – 30 VDC Regulated Power Supply
Calibration Procedure
· With the power supply off, the meter pointer should be over the zero indication on the meter face. If it is not, adjust the mechanical zero-adjust screw at the bottom of the meter until the pointer is properly aligned.
Ancillary Equipment Required
470 ohm 2 watt resistor
* * * * * * *
NOTES
· Remove all jumper wires between the circuit grounds and chassis ground.
· Remove all loads from the supply terminals.
* * * * * * *
Front Panel Initial Conditions
( ) Set the POWER switch to the OFF position.
( ) Set the STANDBY switch to the STANDBY position.
( ) Set the CURRENT control to maximum clockwise (10-turn).
( ) Set the VOLTAGE control to maximum counter-clockwise.
( ) Set the METER switch to the VOLTAGE position
( ) Set the CURRENT RANGE switch to the 0.1 A position.
( ) Set the VOLTAGE RANGE switch to the 30V position.
INITIAL ADJUSTMENTS
( ) Set the HIGH CURRENT ADJ. control (W) [see PICTORIAL 2-7] to maximum counter-clockwise.
( ) Set the POWER switch to the ON position.
( ) Set the STANDBY switch to the DC ON position.
( ) Adjust the VOLTAGE control until the meter reads 30 V.
( ) Set the METER switch to the CURRENT position.
( ) Set the CURRENT switch to the 1 A position.
( ) Set the STANDBY switch to the STANDBY position.
( ) Connect a wire between the + and – output terminals.
( ) Set the STANDBY switch to the DC ON position.
( ) Adjust the HIGH CURRENT ADJ. control (W) until the meter reads 1 ampere full scale.
( ) Set the STANDBY switch to the STANDBY position.
( ) Disconnect the wire between the + and – output terminals.
( ) Set the STANDBY switch to the DC ON position.
OPERATIONAL ADJUSTMENTS
· Figures 5, 7, 8, and 9 depict voltage-current relationships. Voltage is shown on the vertical scale and current is shown on the horizontal scale.
( ) Adjust the VOLTAGE control until the meter reads 20 V.
( ) Set the STANDBY switch to the STANDBY position.
( ) Connect a 470 ohm 2 watts resistor to the output terminals.
( ) Set the STANDBY switch to the DC ON position.
( ) Adjust the CURRENT control counter-clockwise until the voltage just starts to decrease.
· For heavier loads (lower resistance) the current drain will be higher than the example and the limit could be set higher, or the output voltage will drop rapidly to a value appropriate for the current limit selected.
( ) Adjust the VOLTAGE control clockwise. The voltage should not increase.
· The Power Supply is now current limited.
( ) Adjust the VOLTAGE control counter-clockwise until the meter reads less than 20 V.
( ) Adjust the VOLTAGE control clockwise until the meter just reads 20 V.
( ) Set the METER switch to the CURRENT position. The Voltage panel lamp should go out and the amber CURRENT lamp should light.
( ) Adjust the VOLTAGE control clockwise. The current reading should not increase significantly.
( ) Adjust the VOLTAGE control counter-clockwise. The voltage and the current should decrease.
( ) Alternate the METER switch between the CURRENT and VOLTAGE positions.
( ) Set the STANDBY switch to the STANDBY position.
( ) Set the POWER switch to the OFF position.
( ) Disconnect the 470 ohm 2 watts resistor from the output terminals.
( ) Calibration procedure is complete.