Heathkit Model IP-12

Battery Eliminator

 

Date Calibrated:   _____/_____/_____

 

Calibration Procedure

 

·         With the power supply off, both meter pointers should be over the zero indication on each meter face.  If they are not, turn the mechanical zero-adjust screws at the bottom of each meter until the pointers are properly aligned.

 

Ancillary Equipment Needed

 

None

 

* * * * * * *

NOTES

 

·         Remove all loads from the supply terminals.

 

* * * * * * *

INITIAL TEST

 

(    )         Set the POWER switch to the OFF position.

 

(    )         Set the RANGE switch to the 6 VOLTS position.

 

(    )         Adjust the VOLTAGE CONTROL to the maximum counter-clockwise position.

 

(    )         Set the POWER switch to the ON position.

 

(    )         Adjust the VOLTAGE CONTROL slowly clockwise and observe the voltage meter indication.

 

(    )         The meter indication should go from 0 to over 10 volts.

 

·         There should be 0 output indicated on the current meter.

 

(    )         Adjust the VOLTAGE CONTROL slowly to the maximum counter-clockwise position.

 

(    )         The VOLTAGE meter should slowly return to the 0 indication as the charge bleeds off the filter capacitor.

 

(    )         Set the RANGE switch to the 12 VOLTS position.

 

(    )         Adjust the VOLTAGE CONTROL slowly clockwise and observe the voltage meter indication.

 

(    )         The meter indication should go from 0 to over 15 volts.

 

·         There should be 0 output indicated on the current meter.

 

(    )         Adjust the VOLTAGE CONTROL slowly to the maximum counter-clockwise position.

 

(    )         The VOLTAGE meter should slowly return to the 0 indication as the charge bleeds off the filter capacitor.

 

(    )         Set the POWER switch to the OFF position.

 

(    )         Calibration procedure is completed.