Heathkit Model IO-12
Laboratory 5-inch Oscilloscope
Date Calibrated: _____/_____/_____
Calibration Procedure
· Disconnect all external cables, components, and wires from the output terminals.
· All controls and switches are described as labeled on either the instrument or the schematic diagram.
· Use an insulated or non-conducting screwdriver when making all chassis and panel internal adjustments.
Ancillary Equipment Needed
Audio Signal Generator
Frequency Counter
* * * * * * *
FRONT PANEL INITIAL ADJUSTUP
( ) Adjust the INTEN control (R78) to the full counter-clockwise to the OFF position.
( ) Adjust the FOCUS control (R80) to the center of rotation.
( ) Adjust the VERT POS control (R18) to the center of rotation.
( ) Adjust the HOR POS control (R55) to the center of rotation.
( ) Adjust the VERT GAIN control (R8) to the full counter-clockwise position.
( ) Set the HOR / FREQ SELECTOR switch to the EXT INPUT position.
( ) Adjust the HOR GAIN control (R52) to the full counter-clockwise position.
( ) Set the VERT INPUT switch to the X100 position.
( ) Adjust the FREQ VERNIER control (R48) to the 50 position.
( ) Adjust the PHASE control (R72) to the center of rotation.
( ) Adjust the EXT SYNC AMPLITUDE control (R31) to the full counter-clockwise position.
( ) Set the SYNC SELECTOR switch to the EXT position.
( ) Adjust the SPOT SHAPE (R75) [right side of chassis] to the center of rotation.
TRACE ADJUSTMENT
( ) Connect the line cord to the AC power lines.
( ) Adjust the INTEN control (R78) clockwise until the switch clicks to the ON position.
(Allow the instrument to warm up for five minutes.)
( ) Adjust the INTEN control (R78) clockwise until a spot just appears on the display.
( ) Adjust the FOCUS control (R80) to reduce the size of the spot to a minimum.
♦ CAUTION: DO NOT PERMIT A HIGH INTENSITY SPOT TO REMAIN STATIONARY ON THE DISPLAY FOR ANY LENGTH OF TIME. THIS MAY DESTROY THE FLUORESCENT MATERIAL ON THE TUBE AND LEAVE A DARK SPOT.
( ) Adjust the HOR POS control (R55) from full counter-clockwise to the full clockwise position.
(The spot should move horizontally across the display.)
( ) Adjust the HOR POS control (R55) to the center of rotation.
( ) Adjust the VERT POS control (R18) from full counter-clockwise to the full clockwise position.
(The spot should move vertically across the display.)
( ) Adjust the VERT POS control (R18) to the center of rotation.
( ) Adjust the INTEN control (R78) and the FOCUS control (R80) to reduce the size of the spot to a minimum.
( )1 Adjust the SPOT SHAPE (R75) [right side of chassis] to make the spot as round as possible.
( )2 Adjust the INTEN control (R78) and the FOCUS control (R80) to reduce the size of the spot to a minimum.
( ) Repeat the above two steps * for the best spot size and shape.
CATHODE RAY TUBE ALIGNMENT
See Pictorial 9
Construction Manual, page 26
♦ This adjustment only needs to be performed if the trace is not horizontal across the display.
( ) Adjust the INTEN control (R78) to the full counter-clockwise to the OFF position.
( ) Loosen the mounting brackets at the neck of the CRT.
( ) Adjust the INTEN control (R78) clockwise until the switch clicks to the ON position.
(Allow the instrument to warm up for five minutes.)
( ) Adjust the INTEN control (R78) clockwise until a spot just appears on the display.
( ) Connect a test lead to the 1-V P-P terminal.
( ) Connect this same test lead to the HOR INPUT terminal.
( ) Adjust the HOR GAIN control (R52) to the full clockwise position.
( ) Note the trace position on the display.
( )1 Adjust the INTEN control (R78) to the full counter-clockwise to the OFF position.
( )2 Rotate the CRT until the trace should be level.
( )3 Adjust the INTEN control (R78) clockwise until the switch clicks to the ON position.
(Allow the instrument to warm up for five minutes.)
( )4 Adjust the INTEN control (R78) clockwise until a trace just appears on the display.
( ) Repeat the above four steps * until the trace is level.
( ) Adjust the INTEN control (R78) to the full counter-clockwise to the OFF position.
( ) Tighten the mounting brackets at the neck of the CRT.
( ) Adjust the INTEN control (R78) clockwise until the switch clicks to the ON position.
(Allow the instrument to warm up for five minutes.)
( ) Adjust the INTEN control (R78) clockwise until a trace just appears on the display.
( ) Disconnect the test lead from the HOR INPUT terminal.
( ) Connect this same test lead to the VERT INPUT terminal.
( ) Adjust the VERT GAIN control (R8) from full counter-clockwise to the full clockwise position.
( ) Note that the trace is now vertical and controlled by the VERT GAIN control (R8).
( ) Set the VERT INPUT switch to the X10 position.
( ) Adjust the VERT GAIN control (R8) from full counter-clockwise to the full clockwise position.
( ) Note that the trace is now full screen at a lower setting of the VERT GAIN control (R8).
SWEEP GENERATOR OPERATION
( ) Set the SYNC SELECTOR switch to the +INT position.
( ) Adjust the HOR GAIN control (R52) to the 30 position.
( ) Adjust the VERT GAIN control (R8) to the full clockwise position.
( ) Set the HOR / FREQ SELECTOR switch to the dot between the 10 and 100 position.
( ) Adjust the FREQ VERNIER control (R48) for four complete sine waves.
[See Figure 1, Construction Manual, page 31]
· This check indicates that the sweep generator is operating normally at a frequency of 60/4 or 15 Hz.
· The breaks in the trace are caused by the transformer fields. They will not be present with external signals.
( ) Disconnect the test lead from the 1-V P-P terminal.
VERTICAL INPUT TRIMMER ADJUSTMENT
( ) Adjust the INTEN control (R78) to the full counter-clockwise to the OFF position.
( ) Connect this same test lead to the HOR INPUT terminal.
( ) Adjust the INTEN control (R78) clockwise until the switch clicks to the ON position.
(Allow the instrument to warm up for five minutes.)
( ) Adjust the INTEN control (R78) clockwise until a trace just appears on the display.
( ) Set the HOR / FREQ SELECTOR switch to the dot between the 1000 and 10 kHz position.
( ) Adjust the FREQ VERNIER control (R48) to the full counter-clockwise position.
( ) Adjust the INTEN control (R78) clockwise until the switch clicks to the ON position.
(Allow the instrument to warm up for five minutes.)
( ) Adjust the INTEN control (R78) clockwise until a trace just appears on the display.
( ) Adjust both the HOR GAIN control (R52) and the VERT GAIN control (R8) until the diagonal trace is about 2 inches long. [See Figure 2A or 2B, Construction Manual, page 31]
( ) Set the VERT INPUT switch to the X10 position.
( ) Adjust the front trimmer (C1) [left front chassis support] until any trace distortion disappears [Figure 2C].
( ) Set the VERT INPUT switch to the X100 position.
( ) Adjust the rear trimmer (C4) [left front chassis support] until any trace distortion disappears [Figure 2C].
( ) Disconnect the test lead from the VERT INPUT terminal.
( ) Disconnect this same test lead from the HOR INPUT terminal.
( ) Adjust the INTEN control (R78) to the full counter-clockwise to the OFF position.
( ) Disconnect the line cord from the AC power lines.
( ) Assemble the instrument into its case.
ADJUSTING THE PRESET CONTROLS
( ) Connect the line cord to the AC power lines.
( ) Adjust the INTEN control (R78) clockwise until the switch clicks to the ON position.
(Allow the instrument to warm up for five minutes.)
( ) Adjust the INTEN control (R78) clockwise until a trace just appears on the display.
PRESET 1 Adjustment
(10 Hz to 100 Hz)
( ) Connect the Audio Generator output to the Frequency Counter’s input
( ) Adjust the Audio Generator to the desired frequency.
( ) Disconnect the Frequency Counter.
( ) Connect the Audio Generator to the VERT INPUT terminal.
( )1 Set the SYNC SELECTOR switch to the EXT position.
( )2 Set the HOR / FREQ SELECTOR switch to the PRESET 1 position.
( )3 Adjust the VERT GAIN control (R8) for a 1 inch trace.
( )4 Adjust PRESET 1 ADJUST control (R45) [marked hole on front panel] until two complete cycles appear on the display.
( )5 Set the SYNC SELECTOR switch to the +INT position.
( ) Repeat the above five steps * if the trace is not locked.
( ) Disconnect the Audio Generator.
PRESET 2 Adjustment
(1000 Hz to 10 kHz)
( ) Connect the Audio Generator output to the Frequency Counter’s input
( ) Adjust the Audio Generator to the desired frequency.
( ) Disconnect the Frequency Counter.
( ) Connect the Audio Generator to the VERT INPUT terminal.
( )1 Set the SYNC SELECTOR switch to the EXT position.
( )2 Set the HOR / FREQ SELECTOR switch to the PRESET 2 position.
( )3 Adjust the VERT GAIN control (R8) for a 1 inch trace.
( )4 Adjust PRESET 2 ADJUST control (R46) [marked hole on front panel] until two complete cycles appear on the display.
( )5 Set the SYNC SELECTOR switch to the +INT position.
( ) Repeat the above five steps * if the trace is not locked.
( ) Disconnect the Audio Generator.
( ) Adjust the INTEN control (R78) to the full counter-clockwise to the OFF position.
( ) Disconnect the line cord from the AC power lines.
( ) Calibration procedure is complete.