Heathkit Model IM-58
Harmonic Distortion Meter
Calibration Procedure
· With the power supply off, the meter pointer should be over the zero indication on the meter face. If it is not, turn the mechanical zero-adjust screw at the bottom of the meter until the pointer is properly aligned.
· Disconnect all external cables, components, and wires from the input and output terminals.
Ancillary Equipment Needed
AC Voltage Source
AC VTVM
Audio Generator
Frequency Counter
* * * * * * *
INITIAL SETUP
Refer to Pictorial 2
Construction Manual page 21
( ) Disconnect the unit’s line cord from the AC power lines.
( ) Set the CALIBRATE control H [chassis – nearest the meter] to the center of rotation.
( ) Set the COARSE BALANCE control G [chassis – by the power switch] to the center of rotation.
( ) Set the HUM BALANCE control F [chassis – rear center] to full counterclockwise.
( ) Set the RANGE switch [front panel] to the SET LEVEL position.
( ) Set the SENSITIVITY switch [front panel] to the 100% position
( ) Connect the unit’s line cord to the AC power line.
( ) Set the POWER switch [front panel] to the ON position
(Allow the unit to warm up for 30 minutes)
* * * * * * *
VOLTMETER CALIBRATION
Using AC Filament Voltage
( ) Set the SENSITIVITY switch to the 30V position
( ) Connect a test lead to the red INPUT binding post.
( ) Touch the test lead to each one of the pilot light terminals separately.
( ) Connect the test lead to the terminal that gives the highest meter indication.
( ) Set the SENSITIVITY switch to the 10V position.
( ) Adjust the CALIBRATE control H until the meter reads 6.3 volts.
( ) Disconnect the test lead from the INPUT binding post and the pilot lamp terminal.
* * * * * * *
VOLTMETER CALIBRATION
Using a Known AC Source Voltage
( ) Select an AC source voltage close to a full scale meter indication.
( ) Verify the source voltage using an AC VTVM.
( ) Set the SENSITIVITY switch scale position to match the source voltage.
( ) Connect the source voltage to the INPUT binding posts.
( ) Adjust the CALIBRATE control H until the meter indication matches the source voltage.
( ) Disconnect the source voltage from the INPUT binding post.
* * * * * * *
BALANCE ADJUSTMENT
Using an Audio Generator
( ) Set the Audio Generator’s output to 2.0 kHz using a frequency counter.
( ) Set the Audio Generator’s output to 0.5 volts using an AC VTVM.
· Repeat the whole balance adjustment section twice, making adjustments only if necessary the second time.
( ) ( ) Set the SENSITIVITY switch to the 100% position
( ) ( ) Set the RANGE switch to either one of the SET LEVEL positions.
( ) ( ) Connect the Audio Generator’s output to the Distortion Meter’s input terminals.
( ) ( ) Advance the LEVEL control [front panel] to give a full scale meter indication.
( ) ( ) Set the RANGE switch to the 2000–20,000 position.
( ) ( ) Set the BALANCE control [front panel] to the center of its rotation.
( )* ( )* Adjust the TUNING control [front panel] for a minimum meter indication.
( )* ( )* Adjust the COARSE BALANCE control G for minimum meter indication.
( ) ( ) Repeat the above two steps * until no further improvement is obtained.
· It may be necessary to turn the SENSITIVITY switch to a lower percentage value to show the minimum more clearly.
( ) ( ) Set the RANGE switch to the 200–2000 position.
( ) ( ) Set the SENSITIVITY switch to the 100% position
( )* ( )* Adjust the TUNING control for a minimum meter indication.
( )* ( )* Adjust the Tuning trimmer capacitor for a minimum meter indication.
(Through the top hole in the rear chassis shield enclosure [see Figure 15, page 16])
( ) ( ) Repeat the above two steps * until no further improvement is obtained.
· It may be necessary to turn the SENSITIVITY switch to a lower percentage value to show the minimum more clearly.
( ) ( ) Disconnect the Audio Generator from the INPUT binding posts.
( ) ( ) Carefully loosen the setscrew of the insulating coupling [see Figure 10, page 13].
( ) ( ) Adjust the knob pointer to indicate the correct input frequency.
(Make sure the insulated coupling does not turn)
( ) ( ) Carefully retighten the setscrew before continuing.
* * * * * * *
HUM BALANCE ADJUSTMENT
( ) Disconnect all external cables, components, and wires from the input and output terminals.
( ) Disconnect the line cord from the AC power line.
( ) Install the instrument into its case.
( ) Re-connect the line cord to the AC power line.
(POWER switch should still be in the ON position)
( ) Set the RANGE switch to the 20–200 position.
( ) Set the SENSITIVITY switch to the 1% position.
( ) Set the TUNING control to show about 60 Hz on the frequency scale.
( ) Adjust the HUM BALANCE control F to give a minimum indication on the meter.
· The minimum reading should fall around 1/10 of full scale and usually depends upon the 5879 tube’s heater to cathode leakage. Readings above 1/10 full scale may drop below that value after a few hours of continuous operation.
( ) Set the POWER switch to the OFF position.
( ) Disconnect the line cord from the AC power line.
( ) Calibration procedure is complete.