Heathkit Model IM-1212
Digital Multimeter
[Nixie tube model]
Calibration Procedure
· Disconnect all external cables, components, and wires from the output terminals.
Ancillary Equipment Needed
Frequency Counter
* * * * * * *
( ) Disconnect the line cord from the AC power lines.
( ) Set the POWER switch to the OFF position.
* * * * * * *
CIRCUIT BOARD INITIAL ADJUSTMENTS
Refer to Figure 2-2
Construction Manual page 41
( ) Adjust the AC CAL control to the center of rotation.
( ) Adjust the 2M ADJ control to the center of rotation.
( ) Adjust the 200K ADJ control to the center of rotation.
( ) Adjust the 2000 ADJ control to the center of rotation.
( ) Adjust the 200 ADJ control to the center of rotation.
( ) Adjust the ZERO ADJ control to the center of rotation.
( ) Adjust the OSC ADJ control to the center of rotation.
( ) Adjust the AC TP ADJ control to the center of rotation.
( ) Adjust the DC CAL control to maximum counter-clockwise, then a ¼ turn clockwise
* * * * * * *
FRONT INPUT INITIAL ADJUSTMENTS
Refer to Figure 2-1
Construction Manual page 39
( ) Set the FUNCTION switch [SW1] to the OFF position.
( ) Set the SCALE switch [SW2] to the 2 position.
( ) Connect the line cord to the AC power lines.
( ) Set the FUNCTION switch to the DCV position.
(Allow the unit to warm up for 30 minutes)
· The two right-hand digits (displaying any number) and the decimal point to the left should all be lit. No other lamp should be lit.
* * * * * * *
ZERO ADJUSTMENT
Refer to Figure 2-2
Construction Manual page 41
( ) Connect a test lead to the red V input jack.
( ) Connect the same test lead to the black C input jack.
( ) Adjust the ZERO ADJ control for a reading of .01.
( ) Adjust the ZERO ADJ control until the reading just changes to .00.
· The display should change back and forth between .01 and .00.
( ) Disconnect the test lead from both the red V and black C input jacks.
* * * * * * *
OSCILLATOR ADJUSTMENT
(Using a Frequency Counter)
( ) Set the FUNCTION switch to the KW position.
( ) Connect a test lead to the Frequency Counter’s common input.
( ) Connect the same test lead to the Multimeter’s black C input jack.
( ) Connect a test lead to the Frequency Counter’s “hot” input.
( ) Connect the same test lead to either jumper A [see Pictorial 1-2, page 8] or to pin 14 of IC7 (SN7490) [see Pictorial 2-3, page 19].
( ) Adjust the OSC ADJ control for a frequency counter reading of 50 – 60 Hz.
(Match to the known AC line voltage frequency.)
( ) Disconnect both test leads from the Multimeter.
( ) Proceed to the DC VOLTS CALIBRATION section.
* * * * * * *
OSCILLATOR ADJUSTMENT
(Without a Frequency Counter)
( ) Set the FUNCTION switch to the KW position.
( ) Adjust the OSC ADJ control for a reading of OVER .85, within an error of + .15 digits.
· If a reading of OVER .85 is unobtainable, adjust the OSC ADJ control for the nearest possible reading. Then leave the control set at that point.
* * * * * * *
DC VOLTS CALIBRATION
Refer to Figure 2-2
Construction Manual page 41
( ) Connect a test lead to the red V input jack.
( ) Connect the same test lead to the DC CAL TP test point [circuit board – left rear].
· Reference the calibration voltage recorded on the transformer’s insulator to perform the following steps.
· Observe that the calibration voltage number has four digits, whereas the Multimeter can display only three digits. Round the third digit up if the fourth digit is greater than 4.
· If the number is greater than 1.99, the first digit will not be displayed and the OVER lamp shall be lit.
( ) Set the FUNCTION switch to the DCV position.
( ) Adjust the DC CAL control until the reading matches the calibration voltage recorded on the transformer’s insulator.
( ) Disconnect the test lead from both the test point and the red V input jack.
* * * * * * *
OSCILLATOR ADJUSTMENT
(Using a Frequency Counter)
( ) Set the FUNCTION switch to the KW position.
( ) Connect a test lead to the Frequency Counter’s common input.
( ) Connect the same test lead to the Multimeter’s black C input jack.
( ) Connect a test lead to the Frequency Counter’s “hot” input.
( ) Connect the same test lead to either jumper A [see Pictorial 1-2, page 8] or to pin 14 of IC7 (SN7490) [see Pictorial 2-3, page 19].
( ) Adjust the OSC ADJ control for a frequency counter reading of 50 – 60 Hz.
(Match to the known AC line voltage frequency.)
( ) Disconnect both test leads from the Multimeter.
( ) Proceed to the DC CAL Adjustment section.
* * * * * * *
OSCILLATOR ADJUSTMENT
(Without a Frequency Counter)
( ) Set the FUNCTION switch to the KW position.
( ) Adjust the OSC ADJ control for a reading of OVER .85, within an error of +/- .15 digits.
· The reading should fall within .15 digits – no less than OVER .70 or no greater than OVER 1.00.
* * * * * * *
DC CAL ADJUSTMENT
( ) Set the FUNCTION switch to the DCV position.
( ) Connect a test lead to the red V input jack.
( ) Connect the same test lead to the DC CAL TP test point.
( ) Adjust the DC CAL control until the reading matches the calibration voltage recorded on the transformer’s insulator.
( ) Disconnect the test lead from the test point.
* * * * * * *
ZERO ADJUSTMENT
Refer to Figure 2-2
Construction Manual page 41
( ) Connect the same test lead to the black C input jack.
( ) Adjust the ZERO ADJ control for a reading of .01.
( ) Adjust the ZERO ADJ control until the reading just changes to .00.
· The reading should change back and forth between .01 and .00.
( ) Disconnect the test lead from the black C input jack.
( ) Repeat the entire DC VOLTS CALIBRATION section until the correct readings are obtained.
* * * * * * *
AC VOLTS CALIBRATION
Refer to Figure 2-2
Construction Manual page 41
( ) Set the FUNCTION switch to the DCV position.
( ) Set the SCALE switch to the 20 position.
( ) Connect a test lead to the red V input jack.
( ) Connect the same test lead to the AC CAL TP test point [circuit board – left rear, by transformer].
( ) Adjust the AC TP ADJ control for a reading of 16.3.
· The reading may vary between 16.2 and 16.4.
( )* Disconnect the test lead from the test point.
( )* Set the FUNCTION switch to the ACV position.
( )* Connect the same test lead to the AC CAL TP test point, and allow the display to stabilize.
( )* Adjust the AC CAL control for a reading of OVER 0.0.
· The reading may vary between 19.9 and OVER 0.0.
· A proper adjustment will cause it to show a reading of OVER 0.0 most of the time.
( ) Repeat the last four steps * until the correct readings are obtained.
( ) Disconnect the test lead from both the test point and the red V input jack.
* * * * * * *
OHMS CALIBRATION
Refer to Figure 2-2
Construction Manual page 41
( ) Set the FUNCTION switch to the KW (200W) position.
( ) Connect a test lead to the fuseholder [F1] lug 2 [rear].
· The OHMS CAL TP test points are not labeled in Figure 2-2 [circuit board, lift side center].
· All the readings may vary between 19.9 and OVER.
( )1 Set the RANGE switch to the 200 W position.
( ) Connect the same test lead to the 200 W resistor [test point 1 (front)].
( ) Adjust the 200 ADJ control for a reading of OVER 00.
( ) Disconnect the test lead from the test point.
( )2 Set the RANGE switch to the 2 position.
( ) Connect the same test lead to the 2 kW resistor [test point 2].
( ) Adjust the 2000 ADJ control for a reading of OVER .00.
( ) Disconnect the test lead from the test point.
( )3 Set the RANGE switch to the 20 position.
( ) Connect the same test lead to the 20 kW resistor [test point 3].
( ) Adjust the 20K ADJ control for a reading of OVER 0.0.
( ) Disconnect the test lead from the test point.
( )4 Set the RANGE switch to the 200 position.
( ) Connect the same test lead to the 200 kW resistor [test point 4].
( ) Adjust the 200K ADJ control for a reading of OVER 00.
( ) Disconnect the test lead from the test point.
( )5 Set the RANGE switch to the 2K position.
( ) Connect the same test lead to the 2 MW [test point 5].
( ) Adjust the 2M ADJ control for a reading of OVER .00.
( ) Disconnect the test lead from the test point.
( ) Repeat the five ohms adjustments X above until stable readings are obtained.
( ) Disconnect the test lead from the fuseholder [F1] lug 2.
( ) Set the FUNCTION switch to the OFF position.
( ) Disconnect the line cord from the AC power lines.
( ) Calibration procedure is complete.