Heathkit model IG-4244

Oscilloscope Calibrator

 

Date Calibrated:   _____/_____/_____

 

Calibration Procedure

 

•         Turn on the oscilloscope and allow it to warm up and stabilize.

 

•         Disconnect all external components from the front panel terminals.

 

✔  If the instrument is not operating properly, see INITIAL TESTS section. [Construction Manual, pages 49 – 50]

 

Ancillary Equipment Needed

 

Digital Voltmeter (DVM)

25-MHz Oscilloscope

Coax cable [RG58]

Coax cable [RG58] with a 50W load

 

* * * * * * *

INITIAL SETUP

Refer to Pictorial 2-8 Insert

Illustration Booklet, page 7

 

(    )         Set the POWER switch (SW2) to the OFF position.

 

(    )         Adjust the TIME LEVEL control (R2) to the center of rotation.

 

(    )         Set the TIME—SINE—VOLTS switch (SW1) to the VOLTS position.

 

(    )         Install the 2-hole shorting jumper over the two plug block pins (P101) at the DC position [circuit board].

 

(    )         Set the 20 V control (R178) [circuit board] to the center of rotation.

 

(    )         Set the 100 V control (R179) [circuit board] to the center of rotation.

 

(    )         Set the 50 V control (R181) [circuit board] to the center of rotation.

 

(    )         Connect the DVM negative test lead to the outer barrel of the SINE—VOLTS output BNC.

 

(    )         Insert a short length of solid wire or solder into the center conductor of the same BNC.

 

(    )         Connect the DVM positive test lead to the same piece of solid wire or solder.

 

(    )                    Make sure that the heat sink on IC (U118) [circuit board] does not touch the case of transistor (Q113) [Pictorial 2-6, page 6 of the Illustration Booklet]

 

♦  WARNING:  Line voltage is present at several places circuit board.

                          [Pictorial 2-9, page 52 of Construction Manual]

 

(    )         Set the POWER switch (SW2) to the ON position.

 

♦  WARNING:  DO NOT touch the shielded cable center conductor when the RANGE switch (SW101) is in the                      VOLTS position.

 

(    )         Set the RANGE switch (SW101) to the 20 volts position.

 

(    )         Adjust the 20 V control (R178) [circuit board] until the meter indicated exactly 20.00 volts.

 

(    )         Set the RANGE switch (SW101) to the 50 volts position.

 

(    )         Adjust the 50 V control (R181) [circuit board] until the meter indicated exactly 50.00 volts.

 

(    )         Set the RANGE switch (SW101) to the 100 volts position.

 

(    )         Adjust the 100 V control (R179) [circuit board] until the meter indicated exactly 100.0 volts.

 

(    )         Set the RANGE switch (SW101) to the 20 volts position.

 

(    )         Disconnect the DVM test lead from the SINE—VOLTS output BNC.

 

(    )         Remove the wire or piece of solder from the center conductor of the BNC.

* * * * * * *

OSCILLOSCOPE INITIAL SETUP

 

(    )         Set the vertical input attenuator switch to the 5 V/DIV position.

 

(    )         Set the Trigger source switch to INT.

 

(    )         Set the Trigger type switch to AC.

 

(    )         Set the Trigger polarity switch to + (positive).

 

(    )         Set the Trigger mode switch to AUTO.

 

 (    )        Set the Vertical Input attenuator switch to the 0.2 V/cm position.

 

(    )         Set the Horizontal Sweep Rate switch to the .2 msec/cm position.

 

* * * * * * *

TIMING PULSES

Refer to Pictorial 2-10

Construction Manual, page 53

 

(    )         Set the TIME—SINE—VOLTS switch (SW1) to the TIME position.

 

(    )         Set the RANGE switch (SW101) to the 200 nsec position.

 

(    )         Connect the coax cable with the 50W load to the TIME output BNC.

 

(    )         Connect the same coax cable from the Oscilloscope Calibrator to the vertical input of the oscilloscope.

 

* * * * * * *

Refer to Pictorial 2-11

Construction Manual, page 54

 

(    )         Adjust the Oscilloscope Trigger Stability control to obtain a waveform similar to Pictorial 2-11. [Construction Manual, page 54]

 

•         Due to the frequency limitations of many oscilloscopes, the faster rise time waveforms may not be observable.

 

(    )         Set the RANGE switch (SW101) to each frequency position below, and observe the waveform.

 

♦  Readjust the oscilloscope’s horizontal sweep rate as necessary.

 

            (    )   1 s           (    )   500 ms    (    )   200 ms    (    )   100 ms    (    )   50 ms      (    )   20 ms

 

            (    )   10 ms      (    )   5 ms        (    )   2 ms        (    )   1 ms        (    )   500 ms      (    )   200 ms

              

            (    )   100 ms      (    )   50 ms       (    )   20 ms       (    )   10 ms       (    )   5 ms         (    )   2 ms

 

            (    )   1 ms         (    )   500 ns     (    )   200 ns     (    )   100 ns     (    )   50 ns       (    )   20 ns

 

(    )         Disconnect the coax cable with the 50W load from both the oscilloscope and the Oscilloscope Calibrator.

 

* * * * * * *

1-V P-P SINE WAVE

Refer to Pictorial 2-6 Insert

Illustration Booklet, page 6

 

(    )         Set the Oscilloscope Vertical Input attenuator switch to the 0.2 V/cm position.

 

(    )         Set the Oscilloscope Horizontal Sweep Rate switch to the .2 msec/cm position.

 

(    )         Set the TIME—SINE—VOLTS switch (SW1) to the SINE position.

 

(    )         Connect the coax cable without a load to the SINE—VOLTS output BNC.

 

(    )         Connect the same coax cable from the Oscilloscope Calibrator to the vertical input of the oscilloscope.

 

(    )         Adjust control (R192) [bottom side of circuit board] counter-clockwise for the cleanest 1V P-P sine waveform similar to Pictorial 2-12. [Construction Manual, page 55]

 

♦  Adjust the oscilloscope’s vertical attenuator as necessary.

 

(    )         Set the Oscilloscope Calibrator’s POWER switch (SW2) to the OFF position.

 

(    )         Remove the 2-hole shorting jumper from the plug block pins (P101) [circuit board].

 

* * * * * * *

VOLTAGE

Refer to Pictorial 2-13 Insert

Construction Manual, page 56

 

(    )         Set the Oscilloscope Vertical Input attenuator switch to the 0.2 V/cm position.

 

(    )         Set the Oscilloscope Horizontal Sweep Rate switch to the 1 msec/cm position.

 

(    )         Insert the 2-hole shorting jumper over the two plug block pins (P101) at the NORMAL position [circuit board].

 

(    )         Set the TIME—SINE—VOLTS switch (SW1) to the VOLTS position.

 

(    )         Set the RANGE switch (SW101) to the 1 volt position.

 

(    )         Set the POWER switch (SW2) to the ON position.

 

(    )         Adjust the Oscilloscope Trigger Stability control to obtain a waveform similar to Pictorial 2-11. [Construction Manual, page 54]

 

(    )         Set the RANGE switch (SW101) to each voltage position below, and observe the waveform.

 

♦  Adjust the oscilloscope’s vertical attenuator accordingly.

 

            (    )   1 mV       (    )   2mV        (    )   5 mV       (    )   10 mV     (    )   20 mV     (    )   50 mV

 

            (    )   100 mV    (    )   200 mV    (    )   500 mV    (    )   1 V          (    )   2 V          (    )   5 V

              

            (    )   10 V        (    )   20 V        (    )   50 V        (    )   100 V         

 

(    )         Set the RANGE switch (SW101) to the 1 mV position.

 

(    )         Set the Oscilloscope Calibrator’s POWER switch (SW2) to the OFF position.

 

(    )         Disconnect the coax cable from both the oscilloscope and the Oscilloscope Calibrator.

 

(    )         Disconnect the line chord from the line voltage outlet.

 

(    )         Calibration is complete.