Heathkit Model IG-1277
Pulse Generator
Calibration Procedure
· Disconnect all external components from the front panel terminals.
Ancillary Equipment Needed
35 MHz Oscilloscope
50 W cable (RG-58)
50 W Termination Adapter
VTVM
* * * * * * *
FRONT PANEL INITIAL SETUP
Refer to PICTORIAL 4-1
Illustration Booklet, page 4
( ) Set the PERIOD switch [SW101] to the 10 msec position.
( ) Set the DELAY switch [SW102] to the 10 msec position.
( ) Set the WIDTH switch [SW103] to the 1 msec position.
( ) Adjust the PERIOD control [R2] to maximum counter-clockwise.
( ) Adjust the DELAY control [R3] to maximum counter-clockwise.
( ) Adjust the WIDTH control [R4] to maximum counter-clockwise.
( ) Adjust the TRIG LEVEL control [R1] to center of rotation.
( ) Adjust the OFFSET control [R7] to center of rotation.
( ) Adjust the 50W AMPLITUDE control [R5] to center of rotation.
( ) Set the POWER switch to the OFF (out) position.
( ) Set the RUN GATE switch to the ON (in) position.
( ) Set the TRIG DELAY switch to the ON (in) position.
( ) Set all the remaining pushbutton switches to the OFF (out) position.
CIRCUIT BOARD INITIAL SETUP
* * * * * * *
Refer to PICTORIAL 4-2
Illustration Booklet, page 4
( ) Adjust the PERIOD CAL control [R116] to center of rotation [circuit board – right front].
( ) Adjust the WIDTH CAL control [R126] to center of rotation [circuit board – left front].
( ) Adjust the DELAY CAL control [R121] to center of rotation [circuit board – center front].
( ) Adjust the OFFSET CAL control [R151] to center of rotation [circuit board – center rear].
( ) Adjust the Trimmer control [C129] to 9 o’clock position (solder dot) [circuit board – left front].
INITIAL TESTS
* * * * * * *
Refer to Detail 4-2B
Construction Manual, page 40
( ) Set the VTVM FUNCTION switch to the OHMS position.
( ) Set the VTVM RANGE switch to the X1000 position.
( ) Connect the VTVM’s common lead to the heat sink [circuit board].
( ) Connect the VTVM’s probe to one of the line cord’s flat prongs.
( ) VTVM meter indication should be infinite.
( ) Disconnect the VTVM’s probe.
( ) Connect the VTVM’s probe to the other line cord flat prong.
( ) VTVM meter indication should be infinite.
( ) Disconnect the VTVM’s common lead.
( ) Connect the VTVM’s common lead to the line cord’s free flat prong.
( ) Set the POWER switch to the ON (in) position.
( ) VTVM meter indication should be near zero.
( ) Disconnect both the VTVM’s probe and common lead.
( ) Set the POWER switch to the OFF (out) position.
( ) Connect the VTVM’s common lead to the heat sink [circuit board].
( ) Set the VTVM RANGE switch to the X10 position.
( ) Connect the VTVM’s probe to the collector [mounting tab] of transistor Q107.
( ) VTVM meter indication should exceed 25 W.
( ) Disconnect the VTVM’s probe.
( ) Connect the VTVM’s probe to the collector [mounting tab] of transistor Q108.
( ) VTVM meter indication should exceed 25 W.
( ) Disconnect the VTVM’s probe.
( ) Connect the VTVM’s probe to the collector [heat sink] of transistor Q106.
( ) VTVM meter indication should exceed 1000 W.
( ) Disconnect both the VTVM’s probe and common lead.
CALIBRATION
* * * * * * *
Refer to PICTORIAL 4-4
Illustration Booklet, page 15
- - - - - - -
Oscilloscope Setup
( ) Set the oscilloscope’s VERTICLE INPUT to the GND position.
( ) Set the oscilloscope’s VERTICLE INPUT ATTENUATOR to the 0.1 volts/div position.
( ) Set the oscilloscope’s HORIZONTAL SWEEP RATE to the 1 msec/div position.
( ) Adjust the oscilloscope’s trace to the center of the display.
( ) Set the oscilloscope’s VERTICAL INPUT to the DC position.
( ) Connect a 50 W termination adapter to the oscilloscope’s vertical input.
( ) Connect an RG-58 coax cable to the termination adapter.
- - - - - - -
( ) Connect the generator’s line cord to the AC line voltage.
( ) Verify front panel settings are as described in the Front Panel Initial Setup section above.
( ) Connect the same coax cable to the generator’s 50 W OUT jack.
( ) Set the EXT TRIG switch to the ON (in) position.
( ) Set the POWER switch to the ON (in) position.
( ) Adjust the OFFSET CAL control [R151 – circuit board] to center the oscilloscope’s trace.
( ) Switch the oscilloscope’s input between DC and GND several times to verify the trace appears at exactly the same vertical position.
( ) Set the RUN GATE switch to the ON (in) position.
( ) Set the PERIOD switch to the 1 msec position.
( ) Set the oscilloscope’s VERTICAL ATTENUATOR to the 1 volts/div position.
( ) Adjust the oscilloscope’s SWEEP TRIGGER for a trace.
( ) Set the oscilloscope’s HORIZONTAL SWEEP RATE for at least one full visible pulse trace.
( ) Adjust the 50 W AMPLITUDE control to produce a peek-to-peek trace of 5 volts (5 divisions).
( ) Adjust the trimmer capacitor [C129 – circuit board] to produce a slight overshoot on the waveform trace, see Figure A-A [Illustration Booklet, page 15].
· The leading edge of the pulse should appear as straight and vertical as possible.
· Adjustments to both the PERIOD and WIDTH controls may be required.
( ) Set the PERIOD switch to the 10 msec position.
( ) Adjust the PERIOD control to maximum counter-clockwise.
( ) Set the oscilloscope’s HORIZONTAL SWEEP RATE to 2 msec/div.
( ) Adjust the PERIOD CAL control [R116 – circuit board] for a pulse period trace of exactly 5 divisions, see Figure A-B [Illustration Booklet, page 15].
( ) Set the WIDTH switch to the 10 msec position.
( ) Adjust the WIDTH control to maximum counter-clockwise.
( ) Adjust the PERIOD control a quarter-turn clockwise to insure no overlap condition occurs.
( ) Adjust the WIDTH CAL control [R126 – circuit board] for a pulse width trace of 5 divisions, see Figure A-C [Illustration Booklet, page 15].
( ) Disconnect the coax cable from the 50 W termination adapter.
( ) Disconnect the 50 W termination adapter from the oscilloscope’s input.
( ) Connect the coax cable to the oscilloscope’s input.
( ) Disconnect the coax cable from the 50 W OUT jack.
( ) Connect the coax cable to the TRIG OUT jack.
( ) Adjust the DELAY control to maximum counter-clockwise.
( ) Adjust the oscilloscope’s vertical attenuator as necessary.
( ) Adjust the DELAY CAL control [R121 – circuit board] for a trace of 5 divisions wide.
( ) Disconnect the coax cable from the generator.
( ) Disconnect the coax cable from the oscilloscope.
( ) Set the POWER switch to the OFF position.
( ) Disconnect the line cord from the AC power lines.
( ) Calibration procedure is complete.