Heathkit Model EUW-25
DC Oscilloscope
Date Calibrated: _____/_____/_____
Calibration Procedure
♦ The following adjustments were initially all preset at the factory.
· All direction orientations are from the front panel looking aft.
Ancillary Equipment Needed
Audio Signal Generator
Frequency Counter
* * * * * * *
( ) Connect the line cord to the AC power lines.
( ) Adjust the FOCUS control (R56) clockwise until the switch clicks to the ON position.
(Allow the instrument to warm up for five minutes.)
( ) Adjust the FOCUS control (R56) clockwise until the trace is in focus.
* * * * * * *
ASTIGMATISM
( ) Set the VERTICAL switch [rear panel] to the Shorted position.
( ) Set the HORIZONTAL switch [rear panel] to the Shorted position.
♦ CAUTION: DO NOT PERMIT A HIGH INTENSITY SPOT TO REMAIN STATIONARY ON THE DISPLAY FOR ANY LENGTH OF TIME. THIS MAY DESTROY THE FLUORESCENT MATERIAL ON THE TUBE AND LEAVE A DARK SPOT.
( ) Adjust the VERTICAL BEAM control (R69) [left side panel, bottom hole] until the spot is centered vertically on the graticule.
( ) Adjust the HORIZONTAL BEAM control (R76) [right side panel, bottom hole] until the spot is centered horizontally on the graticule.
( ) Adjust the INTENSITY control (R54) [front panel] until the spot is just visible.
( ) ( )1 Adjust the FOCUS control (R56) [front panel] until the spot is in focus.
( ) ( )2 Adjust the ASTIGMATISM control (R51) [rear panel] until the spot is the sharpest.
( ) Repeat the above two * steps as necessary for the smallest and clearest possible spot.
* * * * * * *
VERTICAL DC BALANCE
( ) Set the VERTICAL switch [rear panel] to the Shorted position.
( ) Set the HORIZONTAL switch [rear panel] to the Normal position.
( ) ( )1 Adjust the VERTICAL GAIN control (R8) {red knob} [front panel] until the spot moves.
( ) ( )2 Adjust the VERTICAL DC BALANCE control (R6) [front panel] until the spot is at its original location.
( ) Repeat the above two * steps as necessary until the spot does not move when the VERTICAL GAIN control (R8) is adjusted from maximum counter-clockwise to maximum clockwise.
* * * * * * *
HORIZONTAL DC BALANCE
( ) Set the VERTICAL switch [rear panel] to the Normal position.
( ) Set the HORIZONTAL switch [rear panel] to the Shorted position.
( ) ( )1 Adjust the HORIZINTAL GAIN control (R38) {red knob} [front panel] until the spot moves.
( ) ( )2 Adjust the HORIZONTAL DC BALANCE control (R37) [front panel] until the spot is at its original location.
( ) Repeat the above two * steps as necessary until the spot does not move when the HORIZONTAL GAIN control (R38) is adjusted from maximum counter-clockwise to maximum clockwise.
* * * * * * *
SWEEP CENTERING
( ) Set the VERTICAL switch [rear panel] to the Normal position.
( ) Set the HORIZONTAL switch [rear panel] to the Normal position.
( ) Set the HORIZONTAL ATTENUATOR switch to the AC X 100 position.
( ) Set the FREQUENCY switch to the 5 - 50 position.
( ) Adjust the HORIZINTAL GAIN control (R38) {red knob} [front panel] to maximum counter-clockwise.
· The trace may shift to one side as it becomes a short line.
( ) ( )1 Adjust the SWEEP CENTERING control (R66) [front panel] to position the trace in the center of the display.
( ) ( )2 Adjust the HORIZINTAL GAIN control (R38) {red knob} [front panel] clockwise and observe the trace.
· When centered properly, the trace should expand equally in both directions when the HORIZINTAL GAIN control (R38) is adjusted clockwise.
( ) ( )3 If the trace does not expand equally, adjust the HORIZINTAL GAIN control (R38) {red knob} [front panel] to maximum counter-clockwise.
( ) Repeat the above three * steps as necessary until, the trace is centered and will expand equally.
( ) Set the HORIZONTAL ATTENUATOR switch to the AC X 10 position.
( ) Adjust the HORIZINTAL GAIN control (R38) {red knob} [front panel] to maximum counter-clockwise.
( ) ( )1 Adjust the SWEEP CENTERING control (R66) [front panel] until the trace is centered on the display.
( ) ( )2 Adjust the HORIZINTAL GAIN control (R38) {red knob} [front panel] clockwise and observe the trace.
· When centered properly, the trace should expand equally in both directions when the HORIZINTAL GAIN control {red knob} (R38) [front panel] is adjusted clockwise.
( ) ( )3 If the trace does not expand equally, adjust the HORIZINTAL GAIN control (R38) {red knob} [front panel] to maximum counter-clockwise.
( ) Repeat the above three * steps as necessary until, the trace is centered and will expand equally.
* * * * * * *
CR TUBE POSITIONING
♦ This adjustment only needs to be performed if the trace is not horizontal across the graticule grid.
♦ CAUTION: The voltages on the CRT can be potentially lethal, exercise do care when performing this procedure.
( ) Set the VERTICAL switch [rear panel] to the Normal position.
( ) Set the HORIZONTAL switch [rear panel] to the Normal position.
( ) Adjust the SWEEP GENERATOR control for a horizontal trace.
( ) ( )1 Note the two end points of the trace on the graticule screen.
( ) ( )2 Adjust the FOCUS control (R56) to the full counter-clockwise to the OFF position.
( ) ( )3 Disconnect the line cord from the AC power lines.
( ) Remove the metal cabinet.
( ) Loosen the mounting brackets at the neck of the CRT.
( ) ( )4 Rotate the CRT until the trace should be level on the graticule grid.
♦ Do not allow the CRT to slide forward and come into contact with the graticule grid.
( ) ( )5 Connect the line cord to the AC power lines.
( ) ( )6 Adjust the FOCUS control (R56) clockwise until the switch clicks to the ON position.
(Allow the instrument to warm up for five minutes.)
( ) ( )7 Adjust the FOCUS control (R56) clockwise until a trace is in focus.
( ) If the trace is not horizontal across the graticule grid, repeat ALL the above seven marked * steps.
( ) Adjust the FOCUS control (R56) to the full counter-clockwise to the OFF position.
( ) Disconnect the line cord from the AC power lines.
( ) Carefully retighten the mounting brackets at the neck of the CRT.
♦ Do not allow the CRT to slide forward and come into contact with the graticule grid.
( ) Re-install the metal cabinet.
( ) Connect the line cord to the AC power lines.
( ) Adjust the FOCUS control (R56) clockwise until the switch clicks to the ON position.
(Allow the instrument to warm up for five minutes.)
( ) Adjust the FOCUS control (R56) clockwise until the trace is in focus.
* * * * * * *
VERTICAL ATTENUATOR COMPENSATION
( ) Set the VERTICAL switch [rear panel] to the Normal position.
( ) Set the HORIZONTAL switch [rear panel] to the Normal position.
( ) Adjust a Square Wave generator, using a Frequency Counter, for a 1000 Hz output.
( ) Connect the Square Wave generator output to the [front panel] Vertical input.
( ) Adjust a Sine Wave generator, using a Frequency Counter, for a 1000 Hz output.
( ) Connect the Sine Wave generator output to the [front panel] Horizontal input.
( ) ( )1 Set the VERTICAL ATTENUATOR switch to the AC X 10 position.
( ) ( )2 Adjust the Square Wave generator output for a six graticule peek-to-peek signal.
( ) ( )3 Adjust the X 10 VERTICAL Input trimmer (C3) [left side panel, top front hole] until a correctly compensated square wave is obtained.
( ) ( )4 Set the VERTICLE ATTENUATOR switch to the AC X 100 position.
( ) ( )5 Adjust the Square Wave generator output for a six graticule peek-to-peek signal.
( ) ( )6 Adjust the X 100 VERTICAL Input trimmer (C5) [left side panel, top rear hole] until a correctly compensated square wave is obtained.
( ) Repeat the above six * steps as necessary until, no further trimmer adjustments are required.
( ) Disconnect all inputs from the front panel.
* * * * * * *
HORIZONTAL ATTENUATOR COMPENSATION
( ) Set the VERTICAL switch [rear panel] to the Normal position.
( ) Set the HORIZONTAL switch [rear panel] to the Normal position.
( ) Adjust a Square Wave generator, using a Frequency Counter, for a 1000 Hz output.
( ) Connect the Square Wave generator output to the [front panel] Horizontal input.
( ) Adjust a Sine Wave generator, using a Frequency Counter, for a 1000 Hz output.
( ) Connect the Sine Wave generator output to the [front panel] Vertical input.
( ) ( )1 Set the HORIZONTAL ATTENUATOR switch to the AC X 10 position.
( ) ( )2 Adjust the Square Wave generator output for a six graticule peek-to-peek signal.
( ) ( )3 Adjust the X 10 HORIZONTAL Input trimmer (C15) [right side panel, top front hole] until a correctly compensated square wave is obtained.
( ) ( )4 Set the HORIZONTAL ATTENUATOR switch to the AC X 100 position.
( ) ( )5 Adjust the Square Wave generator output for a six graticule peek-to-peek signal.
( ) ( )6 Adjust the X 100 HORIZONTAL Input trimmer (C17) [right side panel, top rear hole] until a correctly compensated square wave is obtained.
( ) Repeat the above six * steps as necessary until, no further trimmer adjustments are required.
( ) Disconnect all inputs from the front panel.
* * * * * * *
HORIZONTAL BEAM POSITIONING
( ) Adjust the instrument for a trace.
( ) Adjust the HORIZONTAL CENTERING control (R42) until the pointer is centered between the two arrows marked on the front panel.
( ) Adjust the HORIZONTAL BEAM control (R76) [right side panel, bottom hole] until the trace is centered horizontally on the graticule.
* * * * * * *
VERTICAL BEAM POSITIONING
( ) Adjust the instrument for a trace.
( ) Adjust the VERTICAL CENTERING control (R13) until the pointer is centered between the two arrows marked on the front panel.
( ) Adjust the VERTICAL BEAM control (R69) [left side panel, bottom hole] until the beam is centered vertically on the graticule.
* * * * * * *
FINAL ADJUSTMENTS
( ) Set the VERTICAL switch [rear panel] to the Normal position.
( ) Set the HORIZONTAL switch [rear panel] to the Normal position.
( ) Adjust the FOCUS control (R56) to the full counter-clockwise to the OFF position.
( ) Disconnect all wires and test leads from both the front and rear binding posts.
( ) Disconnect the line cord from the AC power lines.
( ) Calibration procedure is complete.