Heathkit model CM-1073
Automotive Tune-up Meter
Date Calibrated: _____/_____/_____
Calibration Procedure
• With the meter off, the meter pointer should be over the 0 volts indication on the meter face. If it is not, adjust the mechanical zero-adjust screw at the bottom of the meter until it does.
• Calibration cable a light duty extension cord with no outlet on the end of it; with a 10K 5% ½ watt resistor (brown-black-orange-gold) soldered to the end of one of the wires. If using a polarized plug extension cord, the resistor should be attached to the hot side of the line (smaller blade of the two).
• CAUTION: DO NOT plug the calibration cable into the 120 VAC 60 Hz line without it being properly connected to the meter test leads first!
• Use a screwdriver with an insulated handle when making the circuit board adjustments. DO NOT come into contact with the metal shaft.
Ancillary Equipment Needed
Automotive battery(s)
Calibration cable
* * * * * * *
INITIAL TEST
Refer to Figure 1
(Fold-out from page 13)
( ) Set the FUNCTION switch to the DC VOLTS position.
( ) The CYLINDERS switch may be in any position.
( ) Connect the black test lead to the negative (--) post of a battery of not over 15-volts.
( ) Touch the red test lead to the positive (+) battery post.
( ) Meter should indicate the voltage of the battery.
( )* If the meter tries to indicate down scale, reverse test leads to the battery.
( ) Disconnect the test leads from the battery.
* * * * * * *
ZERO SET
( ) Position the instrument so that the meter is facing upward.
( ) Connect (short) the two test leads together.
( ) Adjust the meter Zero Set screw so that the meter pointer reads exactly zero.
( ) Disconnect the two test leads from each other.
* * * * * * *
DWELL Calibration
Refer to Figure 2
Note: If possible, this calibration should be performed using a standard fully charged 6-volt automotive battery. After calibration, the instrument should be checked using a fully charged 12-volt automotive battery. The meter readings should be approximately the same for both automotive voltages. If a 6-volt battery is unavailable (older automotive electrical systems), a standard 12-volt battery will work just fine.
( ) Set the FUNCTION switch to the DWELL position.
( ) Set the CYLINDERS switch to any position.
( ) Observing correct polarity, connect the black test clip to the negative (-) terminal and the red test clip to the positive (+) terminal of any DC voltage source between 5 and 15 volts, such as an automotive battery.
( ) Adjust the DWELL CAL control [located near the top of the circuit board] until the meter reads full scale.
( ) Disconnect the test clips from the voltage source.
* * * * * * *
TACHOMETER CALIBRATION
Refer to Figure 3.
• Use extreme care when using AC line voltage. Personal shock or meter damage can result from carelessness.
( ) Set the FUNCTION switch to the LOW RPM position.
( ) Set the CYLINDERS switch to the 6 position.
( ) Set the TAC CAL control [located near the bottom of the circuit board] to midrange.
( ) Connect the meters test clips to the calibration cable: [red alligator clip to the resistor lead - the resistor needs to be in series with the meter], and the [black alligator clip to the plain lead].
• The calibration cable should not remain plugged into the AC line any longer than necessary to accomplish the calibration (about 2 3 minutes). The resistor will overheat if left on for too long a time.
• CAUTION: Use a screwdriver with an insulated handle.
( ) Plug the calibration cable into a 120 VAC 60 Hz outlet.
( ) Adjust the TAC CAL control until the meter reads 1200 RPMs (12 on the 0 15 scale).
( ) Set the CYLINDERS switch to the 8 position.
( ) The meter should read 900 RPMs (9 on the 0 15 scale).
( ) Set the FUNCTION switch to the HIGH RPM position.
( ) Set the CYLINDERS switch to the 4 position.
( ) The meter should read 1800 RPMs (18 on the 0 45 scale).
( ) Set the CYLINDERS switch to the 3 position.
( ) The meter should read 2400 RPMs (24 on the 0 45 scale).
• CAUTION: Always unplug the calibration cable from the AC line first!
( ) Disconnect the test cable from the 120 VAC 60 Hz outlet.
( ) Disconnect the test leads from the calibration cable.
( ) Calibration procedure is complete.