IT-2240 DIGITAL LC BRIDGE The Heathkit IT-2240 Digital LC Bridge is designed for easy measurement of inductance and capacitance, and their associated loss (dissipation) factor. Its accuracy and operational features can fulfill the requirements of a laboratory, a quality assurance department, or an experimenter. The clearly labeled front panel lets even untrained users make error-free measurements. SPECIFICATIONS Accuracy specifications apply at 22C (72F) after a 10 minute warm-up. L (INDUCTANCE) Ranges. . . . . . . . . . . . . . . . . . 200 uH to 2000 H; 8 decades Accuracy (+/- 4 counts)*. . . . . . . . . +/- [.5% (L) + [1.5% (L) x D + .5 ?H] for readings from 10 to 2000 counts when the Bridge is calibrated using laboratory standards Excitation. . . . . . . . . . . . . . . . 1000 Hz sinusoidal: 80 mA on 200 uH range decading to 8 ?A on 2 H range. 120 Hz sinusoidal: 6 uA on 20 uH range; .6 uA on 200 H, 2000 H ranges. Impedance Model . . . . . . . . . . . . . Series R, L equivalent C (CAPACITANCE) Ranges. . . . . . . . . . . . . . . . . . 200 pF to 2000 uF; 8 decades Accuracy (+/- 2 counts)*. . . . . . . . . +/- [.5% (C) + [1.5% (C) x D + .5 pF] for readings from 10 to 2000 counts when the Bridge is calibrated using laboratory standards Excitation. . . . . . . . . . . . . . . . 1000 Hz sinusoidal: .8 VAC on 200 pF to 2 uF ranges Impedance Model . . . . . . . . . . . . . Parallel R, C equivalent DC Bias . . . . . . . . . . . . . . . . . +/- .01 VDC to 10 VDC may be applied via rear terminals. Up to 50 VDC may be applied using eternal circuitry. D (DISSIPATION) Ranges. . . . . . . . . . . . . . . . . . 000 to 1.999, limited to D x uF = 1000 on 2000 uF range Accuracy (+/- 20 counts)* . . . . . . . . +/- 3% for value displays greater than 200 counts GENERAL Measurement Connection. . . . . . . . . . Front panel slotted inputs accommodate leads up to 1/16” in diameter and 2-1/2” apart, and provide 4-terminal (Kelvin) measurement. Accessory assembly (provided) allows 4-terminal measurements of large or remote parts. Excitation Frequency Accuracy . . . . . . +/- 3% Overvoltage Protection. . . . . . . . . . Input protected against +/- 20 VDC/VAC (5 seconds maximum), or capacitor up to 2000 ?F charged to 50 VDC. Higher overvoltage will cause an internal 1-ampere fuse to open. Display . . . . . . . . . . . . . . . . . 3-1/2 digit (1999 maximum count), .43” high-brightness LED Sampling Rate . . . . . . . . . . . . . . 2-1/2 per second Value Out . . . . . . . . . . . . . . . . 0 to +2.0 VDC (+/- 5%), proportional to L, C value, provided at rear panel terminals. Temperature Coefficient . . . . . . . . . +/- (.1 x Accuracy)/C for L, C +/- (.05 x Accuracy)/oC for D Operating Temperature . . . . . . . . . . 10 to 40C (50 to 104F) Storage Temperature . . . . . . . . . . . 0 to 50C (32 to 122F) Power Requirements. . . . . . . . . . . . 105 - 125 VAC or 210 – 250 VAC, 50/60 Hz, 15 Watts Dimensions. . . . . . . . . . . . . . . . 3.75” high x 10” wide x 12.5” deep Net Weight. . . . . . . . . . . . . . . . 6.5 lbs