Heathkit IM-11 VTVM Meter scales: DC & AC (RMS): 0-1.5, 5, 15, 50, 150, 500, 1500 volts full scale AC peak-to-peak: 0-4, 14, 40, 140, 400, 1400, 4000 Resistance: 10 ohm center scale at x1, x10, x100, x1000, x10K, x100K, x1 meg Measures .1 ohm to 1000 megohms with internal battery Meter: 4½" 200 ua movement Multipliers: 1% precision type Input resistance DC: 11 megohms (1 megohm in probe) on all ranges Circuit: Balanced bridge (push pull) using twin triode Accuracy: DC ± 3%, AC ± 5% at full scale Frequency response: ± 1 db, 25 cps to 1 mc (600 ohm source) Tubes: 12AU7, 6AL5 Battery: 1.5 volt, size "C" flashlight cell Power requirements: 105-125 volt 50/60 cycle AC 10 watts Dimensions: 7 3/8" H x 4 11/16" W x 4 1/8" D