Heathkit Model AD-1309
Pink/White Noise Generator
Calibration Procedure
· Set the POWER OUTPUT switch (on front of unit) full counter-clockwise to the OFF position.
· Disconnect all external components from the front panel terminals.
· Install a fresh 9-volt battery
· Calibration is required only when used with a real-time spectrum analyzer.
Ancillary Equipment Needed
Spectrum Analyzer
Shielded cable
* * * * * * *
HIGH FREQUENCY ADJUSTMENT
Refer to PICTORIAL 4-1
Construction manual page 20
( ) Set the spectrum analyzer power switch to the ON position and allow the unit time to stabilize.
( ) Set the spectrum analyzer input attenuator to the 1 dB/div scale.
( ) Connect a shielded cable to the external input jack of the spectrum analyzer.
( ) Connect the shielded cable to the generator’s pink noise output jack J2.
( ) Set the generator’s POWER OUTPUT control to the ON position.
( ) Adjust the POWER OUTPUT control clockwise until a meaningful display is shown on the spectrum analyzer.
( ) Adjust the spectrum analyzer external input attenuator control as necessary.
· If the spectrum analyzer display shows a flat response from 1 kHz to 20 kHz – no calibration is needed.
· If a .0022 uF capacitor is installed into the two sockets at C3 on the circuit board, adjust trimmer R1 as necessary to obtain the flattest frequency response.
· If there is unacceptable high-frequency roll off with no capacitor at C3 – continue as follows:
( ) Insert a .0022 uF Mylar capacitor into the two sockets labeled C3 next to the 10 uF electrolytic capacitor C4 on the circuit board.
( ) Adjust the trimmer R1 on the circuit board through its range until the flattest frequency response is displayed on the spectrum analyzer.
· If the frequency response obtained using C3 is NOT an improvement or too much high-frequency boost is generated – remove capacitor C3. Calibration is now completed. Go to the last five steps marked * below.
· If the frequency response obtained using C3 is an improvement – solder capacitor C3 into the two sockets.
( ) Adjust the trimmer R1 on the circuit board through its range until the flattest frequency response is displayed on the spectrum analyzer.
( )* Set the generator’s POWER OUTPUT control to the OFF position.
( )* Disconnect the shielded cable from the generator’s pink noise output jack J2.
( )* Disconnect the shielded cable from the external input jack of the spectrum analyzer.
( )* Set the spectrum analyzer power switch to the OFF position.
( )* Calibration procedure is complete.